OMEGA FP-5060 Series Micro Flow Sensor

Technical Notes:
• Horizontal surface mount (±30 °) recommended for optimum performance.
• Suitable for clean fl uids only. Suspended solids will cause mechanical failure.
• ¼ in. NPT or ISO 7/1 - R¼ (male) port connections
• NEMA 4X/IP65 splashproof enclosure

30°

30°

Recommended

mounting angles

Vertical mounting

not recommended

OMEGA FP-5060 Series Micro-Flow Sensors

2. Recommended Sensor Position

1. Instrument

Wiring

Technical Notes:

Use 2-conductor twisted-pair shielded cable for cable splices up to 300 m (1000 ft) max.

Maintain cable shield through cable splice.

Route sensor cable away from AC power lines.

AUX power required when used with any OMEGA FP90 Series Flow Transmitter.

OMEGA FP-5060
Series Sensor

Input

Gnd.

10 kΩ

+

5 to 24
VDC

-

Black

Shield

Red

Other
instrument

Black (5 to 24 VDC)

Red (Signal out)

Shield (DC return)

Blk
Power

Red
Freq. input
Shld
Gnd

Other

Instrument

OMEGA FP-90 Series, FPM-5500,

FPM-9020A

Note: A

10 kΩ pull-up resistor is required as shown, if

not provided internal to the instrument. Contact
OMEGA Technical Support for assistance.

SAFETY INSTRUCTIONS
1.

Do not remove from pressurized lines.

2.

Do not exceed maximum temperature/pressure specifi cations.

3.

Do not install/service without following installation instructions (see sensor manual).

4.

Wear safety goggles and faceshield during installation/service.

5.

Do not alter product construction.

6.

Failure to follow safety instructions could result in severe personal injury!