3. Calibration


The OME-A-8111 is calibrated to its best state of operation. For environments with large
vibration, recalibration is recommended. Before calibrating the OME-A-8111, users should
have the following items:

z

One 6-digit multi-meter.

z

One stable voltage source (4.9988V)

z

Diagnostic program: this program is included in the delivered package

and will guide the user to in proceeding with the calibration.




3.1 Calibration VR Description


There are seven VRs on the OME-A-8111. Calibration is needed to adjust all seven VRs.

VR Num. Description

VR1

D/A Gain adjustment

VR2

D/A Offset adjustment

VR3

A/D Offset adjustment

VR4

A/D's Gain adjustment

VR5

A/D‘s PGA(Programmable Gain Amplifier) Offset Adjustment

VR4

A/D Converter

VR3

D/A Converter

PGA

VR5

VR1

O.P.

VR2

CN1

8 ch.
Mux










OME-A-8111 Hardware Manual (ver.1.1, Jul/2003) 29