8

4.2 Programming Flowchart

RATE

RA

RA

TOTAL

CLR ToT

FDEC

12345678

M

RUN MODE

CLEAR TOTAL

OT

OT

FACTOR
DECIMAL

FAC

100

FACTOR

0000

ent Code

ENTER CODE

If panel
lock ON

If panel
lock OFF

If code
correct

If code
incorrect

FAC

20poInt

FACTOR
TYPE

20poInt

lInear

lInear

selected

E

M

E

E

M

E

M

E

M

20poInt

selected

Continue On

Next Page

Fr#

#####

FREQUENCY
for POINT #
(Fr0-Fr19)

FaC#

#####

FACTOR for
POINT #
(FAC0-FAC19)

If Fr1-FR19 = 0

# = 19

E

M

E

M

YES

SET pnts

E

M

YES

No

SET

20PT

YES

selected

No

selected

Press the M key to enter the programming menu.

The

SET pnts prompt will only appear if 20point is selected. This allows users

to bypass the 20 point set up and keep the existing values.

Press the key to choose YES or NO.
Press the E key to enter the displayedselection.
Press the M key to skip (same as selecting NO).

If the panel lock is on, you must enter the 4 digit lock code to gain access

to the menu.

Press the key to increment each individual digit of the code.
Press the

key to advance to the next digit.

Press the E key to enter the displayed code.
If the code is correct, display advances to "

clr tot", if not, display returns

to run mode

Press the E key to clear the totalizer and return to the programming menu.
Press the M key to skip and go to next menu item.

Press the

key to step the factor decimal to the desired location.

Press the E key to enter the displayed decimal location.
Press the M key to skip and keep the existing location

NOTE: For best performance and resolution choose as many decimal places as
possible in the K-Factor.Example: Enter a K-Factor of 1 as 1.000.

Press the key to choose factor type (

20point or linear).

Press the E key to enter the displayed factor type.
Press the M key to skip and keep the existing factor type.

Press the key to increment each individual digit of the factor for point #.
Press the

key to advance to the next digit.

Press the E key to enter the displayed factor.
Press the M key to skip and keep the existing factor.
If 0 is entered, the display will advance to the next prompt (tdec).

Press the key to increment each individual digit of the frequency for point

#.

Press the

key to advance to the next digit.

Press the E key to enter the displayed frequency.
Press the M key to skip and keep the existing frequency.
If 0 is entered, the display will advance to the next prompt (tdec).

NOTE: Frequency/factor point pairs must be entered in ascending order of fre-

quency

The

FAC prompt will only appear if linear is selected or if the unit was ordered

without the 20 point linearization option.

Press the key to increment each individual digit of the factor.
Press the

key to advance to the next digit.

Press the E key to enter the displayed factor.
Press the M key to skip and keep the existing factor.