24

8.0 Default Alarm and Configurations

Table 3: Default Alarm

Gas

Full Scale

Display

Resolution

Alarm 1

Alarm 2

TWA

STEL

Calibration

Gas Value

Ammonia

99 ppm

1 ppm

25 ppm

50 ppm

25 ppm

35 ppm

25 ppm NH

3

Arsine

1.00 ppm

0.01 ppm

0.05 ppm

0.10 ppm

0.10 ppm

0.10 ppm

0.5 ppm PH

3

0.5 ppm /AsH

3

BRH

200 ppm

1 ppm

100 ppm

150 ppm

X

X

100 ppm C

7

H

8

(toluene)

Carbon monoxide

500 ppm

1 ppm

35 ppm

200 ppm

35 ppm

200 ppm

100 ppm CO

Chlorine

9.9 ppm

0.1 ppm

0.5 ppm

5.0 ppm

.05 ppm

1.0 ppm

5 ppm Cl

2

Chlorine dioxide

3.00 ppm

0.01 ppm

0.10 ppm

0.30 ppm

0.10 ppm

0.30 ppm

1.0 ppm Cl

2

CO+ (BRH)

100 ppm

1 ppm

35 ppm

75 ppm

X

X

35 ppm CO

Combustible

99 %LEL

1 %LEL

10 %LEL

50 %LEL

X

X

50 %LEL CH

4

HS+ (BRH)

40 ppm

1 ppm

10 ppm

20 ppm

X

X

20 ppm H

2

S =

50 ppm CO

Hydrogen chloride

20.0 ppm

0.1 ppm

5.0 ppm

10.0 ppm

5.0 ppm

5.0 ppm

10 ppm HCl

Hydrogen cyanide

20.0 ppm

0.1 ppm

4.7 ppm

10.0 ppm

4.7 ppm

4.7 ppm

10 ppm HCN

Hydrogen fluoride

10.0 ppm

0.1 ppm

0.3 ppm

6.0 ppm

3.0 ppm

3.0 ppm

5.0 ppm HF

Hydrogen sulfide

200 ppm

1 ppm

10 ppm

20 ppm

10 ppm

15 ppm

20 ppm H

2

S

Ozone

3.00 ppm

0.01 ppm

0.10 ppm

0.30 ppm

0.10 ppm

0.10 ppm

0.3 ppm O

3

Oxygen

30 %volume

0.1 % volume

19.5%

23.5%

X

X

20.9 % O

2

Phosphine

2.00 ppm

0.01 ppm

0.30 ppm

1.50 ppm

0.30 ppm

1.5 ppm

0.5 ppm PH

3

Nitrogen dioxide

30.0 ppm

0.1 ppm

1.0 ppm

5.0 ppm

1.0 ppm

1.0 ppm

5 ppm NO

2

Sulfur dioxide

30.0 ppm

0.1 ppm

2.0 ppm

10.0 ppm

2.0 ppm

5.0 ppm

10 ppm SO

2

Ethylene oxide

10 ppm

0.1 ppm

5.0 ppm

7.5 ppm

5.0 ppm

5.0 ppm

5 ppm ETO

X = not applicable.

These gases require the Sampling Shield p/n 04836-000 to be in place and pump activated for

accurate detection and calibration.

Sensor response time with internal pump, if the tubing is 20 feet long, allow at least 10 seconds for the test sample
to make it through the tubing. The sensor response time is typically 30 – 60 seconds to achieve 90% of the final
reading, depending on the specific sensor and the gas being detected. Therefore, the total 'system' response time for an
instrument using the internal sampling pump and 20 feet of hose is approximately 40-70 seconds to achieve 90% of the
final reading.

Table 4: Default Configuration

Operation Menu:

Basic

Confidence chirp interval:

120 seconds

Backlight duration:

30 seconds

Combustible gas display:

Methane

Datalogging:

Active (requires

TARGET

Trakker to retrieve)

Datalogging interval:

1 per minute

TWA/STEL Alarms:

Active

Locations:

Enabled (requires

TARGET

Trakker to set)