ALL VERSIONS WITH 0-5 VDC CIRCUIT

Power requirements:

12-35VDC, 20 mA

Maximum Current:

30 mA DC

Max. Trans. Distance:

<200 feet recommended

Min. Load Resistance:

1000

Flow rate accuracy:

2% of full scale

Repeatability:

0.5% of scale

Resolution:

infinite

Response time:

2 seconds to 90%

ALL VERSIONS WITH PULSE OUTPUT CIRCUIT

Power requirements:

5-24VDC, 20 mA

Maximum Current:

25 mA DC

Max. Trans. Distance:

<200 feet recommended

Flow rate accuracy:

2% of full scale

Linearity:

0.5% of scale

Resolution:

Infinite

Response time:

< 100 mS

ALL VERSIONS WITH RELAY OUTPUT CIRCUIT

Power Requirements

12-35 V

DC

Max. Trans. Distance

< 200 feet recommended

Switch Contact

Form C, 5A max @120 or 240 VAC

Hysteresis

5% of reading max.