Omega Vehicle Security DP97 User Manual

Page 17

Advertising
background image





b) Using the software

From MAIN MENU:-

To input calibration values into the DP97 (up to 10 may be used per channel):

º

Click EDIT to obtain EDIT corrections menu.

º

Enter values against #01 through #10 but note that


i)

All 10 need not be entered

ii)

When returning to main menu the Ù values are sorted into ascending order.
If the corresponding °C values are not in ascending order, the status box
will show Error.

iii)

All values must be entered with up to three decimal places e.g. 0.001,
10.001 etc. Both °C and Ù values must be entered.

º

Click return to Main Menu

º

From Main Menu, click FILES to obtain Load/Save/Print Menu

º

Click SPECIFY

º

Key file name into ‘CURRENT SELECTION’ File names end with .COR eg.
EXPI.COR. If .COR is not keyed in, it will be added automatically when ENTER
is pressed or OK clicked.

Note: File name should be cross referred to probe serial number.

º

Click SAVE to save file

º

Click LOAD to load data

º

Click RETURN TO MAIN MENU and note loaded values are shown under
“Calibration Points”

º

Click CHANNEL A <-> B

º

Click SEND to transmit correction values to DP97

º

Click CORRECT A, B OR A + B to switch on corrections


Note: Red dot in lower right hand corner of DP97 display window lights during “comms”

activity.

º

Observe Channel A/B status LED on DP97 occults to indicate corrections stored
Within defined range. If this does not occult, probe reading does not lie within
corrections range and corrections are not used.