P6701B, P6703B, and P6723 Service Manual

7

P6701B and P6703B Performance Verification

Use the following procedures to verify the warranted specifications of the
P6701B and P6703B O/E converters. Before beginning these procedures, refer to
page 12 and photocopy the test record and use it to record the performance test
results. The recommended calibration interval is one year.

These procedures test the following specifications:

H Noise equivalent power

H Output zero

H DC conversion gain

H AC conversion gain

Equipment Required

Table 7 lists the equipment required to perform the performance verification
procedure. The types and quantities of connectors may vary depending on the
specific equipment you use.

The instrument should be given a 20 minute warm up period and ambient
temperature must be between 20 and 30 degrees C.

Table 7: Test equipment

Description

Minimum requirements

Example product

Optical power meter with
head and adapters

Accuracy > 2.5%, resolution >
5 pW, Max power > 1 mW, cali-
brated from 700 nm -- 1600 nm

HP 8153A with power
sensor HP 81532A and
HP 81531A

P6701B only: 780 nm cal
source

output > 200uW (CW)

1

,

stability > 0.1 dB over 5 minutes,
modulated square wave @ 10 kHz
with off modulation at zero-light
level

P6701B only: 850 nm cal
source

output > 200uW

1

,

stability > 0.1 dB over 5 minutes,
modulated square wave @ 10 kHz
with off modulation at zero-light
level