P6701B and P6703B Performance Verification

P6701B, P6703B, and P6723 Service Manual

11

3. Set the oscilloscope for 5 us/div on the horizontal scale and 5 mV/div (or

5 uW/div) for the vertical scale. Trigger the scope on the negative edge of
the trace and set ground to the middle of the screen. This is equivalent to
1%/div.

4. Check that the OFF level of the square wave at approximately 5 us after the

90% fall time point is within 16 mV (8%) of the ground level (center line).