P6723 performance verification, Equipment required – Tektronix P6723 User Manual

Page 23

Advertising
background image

P6701B, P6703B, and P6723 Service Manual

13

P6723 Performance Verification

Use the following procedure to verify the warranted specification (output
amplitude) of the P6723 O/E converter. Before beginning these procedures, refer
to page 15 and photocopy the test record and use it to record the performance test
results. The recommended calibration interval is one year.

Equipment Required

Table 8 lists the equipment required to perform the performance verification
procedure. The types and quantities of connectors may vary depending on the
specific equipment you use.

The instrument should be tested at an ambient temperature between 20 and 30
degrees C.

Table 8: Test equipment

Description

Minimum requirements

Example product

Optical power meter with
head and adapters

Accuracy > 2.5%, resolution >
5 pW, Max power > 1 mW, cali-
brated from 700 nm -- 1600 nm

HP 8153A with power
sensor HP 81530A and
HP 81531A

1310 nm digital optical
signal source

52 MHz -- 622 MHz

Tektronix CTS710/CTS750

Sampling oscilloscope

oscilloscope with Tekprobe II
interface

Tektronix TDS784x

Adjustable optical
attenuator

4 decades, 50 m core fiber,
FC-style connectors

JDS 5000L with 62 m
fiber with FC connectors

Optical cable

FC-FC multimode, 62.5 m, 2
meters

174-2322-00