The specific descriptions for each type of alarm are as follows:

Power Fail Alarm: This alarm changes state when the AC or DC
system power is lost.

Alarm #1: Programmable as high or low alarm (actuates when
concentration is above threshold). Can be set anywhere between 0
and 9999ppm, but must be set ABOVE the threshold set for Alarm 2.

Alarm #2: Programmable as high or low alarm (actuates when
concentration is below threshold). Can be set anywhere from 0 to
9999ppm.

CAUTION:

THERE COULD BE HAZARDOUS VOLTAGE AT THE
ALARMS TERMINALS, EVEN WHEN POWER IS
REMOVED FROM THE INSTRUMENT.

3.3.4 Relay Ratings

The 3300/OT-3 has 2 relay ratings.

Higher rating —established by the manufacturer

DC Rating:

0-30 VDC maximum @ 2.0 Amps
(resistive)

AC Rating:

125 VAC maximum @ 0.5 Amps

minimum reliable switching rating:

10

µ

A @ 10mV

Lower rating—maximum safe rating for Cl 1 Div 2 Group
C or D hazardous location

DC Rating:

0-30 VDC maximum @ 0.9 Amps
maximum

AC Rating:

125 VAC maximum @ 0.2 Amps
maximum

minimum reliable switching rating:

10

µ

A @ 10mV

To safely use the alarm relays in a hazardous location, the

characteristics of both the switched voltage and the components or load
being operated must be known. The alarm relays must not be attached to