Teledyne OT-3 User Manual

Page 33

Advertising
background image

Trace Oxygen Analyzer

Operation

Teledyne Analytical Instruments

25

4.3.1 HI Range

Setting the HI Range full-scale value defines the LEAST sensitive

analysis range to be used. To set the HI Range:

1. Press the SET HI RANGE Function button once.

2. Immediately (within 5 seconds) press either the

or

button to

raise or lower the displayed value, as required, until the display
reads the desired full-scale percent concentration.

4.3.2 LO Range

Setting the LO Range full-scale value defines the MOST sensitive

range to be used. To set the LO Range:

1. Press the SET LO RANGE Function button once.

2. Immediately (within 5 seconds) press either the UP or DOWN

arrow button to raise or lower the displayed value, as required,
until the display reads the desired full-scale percent
concentration.

4.4 Setting the Alarm Setpoints

The alarm setpoints can be adjusted over the full range of the

analyzer (0-9999 ppm oxygen content). The set point values are
expressed in ppm only.

4.4.1 Set Alarm 1

Alarm 1 can be set either as a high or low alarm. To configure this

alarm to your preferences:

1. Press the SET ALARM 1 function button once.

2. The display will flash either HI or LO depending on what the

alarm was configured in the last time. If the alarm configuration
must be changed, use the Up or Down keys to toggle the alarm
from HI to LO or vice-versa (if within 5 seconds no key is
pressed, the instrument will return to the sample mode and
display oxygen concentration).

3. To change the setting at which the alarm will be actuated, press

the SET ALARM 1 function button once more. The alarm
setpoint will flash on the LED display. Press either the Up or
Down keys to raise or lower the displayed value, as required,

Advertising
This manual is related to the following products: