Appendix – Teledyne OT-3 User Manual

Page 44

Advertising
background image

Appendix

OT-3

Teledyne Analytical Instruments

36

Appendix

A.1

Specifications

Ranges: Two user selectable ranges can be set

between 0-10 ppm and 0-9999 ppm
oxygen, and a 0-25 % (nominal) Air
Calibration Range.

Signal Output: Voltage:

0–10 VDC, negative ground

Current:

4-20 mA, negative ground

Range ID: 0-10 VDC.

Display: Light emitting diode (LED) display.

Alarms: Two user-settable non-latching alarms,

with user adjustable programmable 0-20
min.delay; one power failure relay. (All
are failsafe.)

System Operating Temp: 0-50 °C

Accuracy: ±2 % of full scale at constant temperature

±5 % of full scale through operating
temperature range (At 100 ppm and higher
user defined ranges) once temperature
equilibrium is reached.
±1 ppm for 10 ppm range under above
conditions.

Response Time: 90 % in less than 65 seconds at 25°C (68°F).

System Power Requirement:

AC (100-240 VAC, 50/60Hz @ 2.8W), or
DC (10-36 VDC @2.8W )

Weight: 30.9Lbs (14 Kg)

System Enclosure: NEMA 4 enclosure.

40 cm x 30 cm x 20.6 cm
(15.8” x 11.8” x 8.1”)

Advertising
This manual is related to the following products: