; Read the Operators Manual thoroughly before using the instrument.

; Calibrate once per shift, and prior to each and every use.

; Check the high and low alarm threshold settings prior to use.

; Replace the batteries immediately when required.

; Keep the instrument, cable and sensor dry at all times.

; Re-calibrate after changing the sensor or batteries.

; Visually inspect the whole instrument prior to use.

; Check the integrity of all cable connections routinely.

; Remove and carefully store the “flow diverter” when using the instrument in “static”

environments such as incubators, oxygen hoods etc.

; Perform an alarm functionality test prior to each use.

; Remove batteries if the instrument is not to be used for a month plus

D

ON

TS

: Use the Accurox Oxygen Monitor if you suspect any malfunction.

: Use the monitor in the presence of flammable anaesthetics.

: Autoclave the instrument and/or the sensor.

: Submerge the instrument and/or sensor in water or any other liquid.

: Expose instrument to other equipment with high levels of RFI emission.

: Place the instrument itself in a water vapour saturated environment.

: Expose the LCD to excessive direct sunlight

: Dispose of the sensor cell in fire, or place in an incinerator.