20

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

S

ECTION

S

EVEN

T

ECHNICAL

S

PECIFICATIONS

SAFETY
CLASSIFICATION

Complies to IEC 601 – Safety of Medical Equipment Internally
powered equipment

Drip proof – IPX 1

Not suitable for used with flammable anaesthetic mixture

POWER

Internal : DC 6V, 4 dry cell batteries – size AA/penlight

MASS

525 grams (with T-7 oxygen sensor)

DIMENSIONS

59mm (H) x 152mm (W) x 115mm (D)

BATTERY DURATION

Approximately 900 hours of continuous non-alarming operation

Low battery status indication is provided

Replace batteries within 48 hours of low battery indication

ALARMS

High oxygen concentration (18% to 102%)

Low oxygen concentration (18% to 100%)

Low battery (visual alarm only)

Most alarm conditions are signalled audibly and visually, with a 60
second audible alarm mute facility provided

MEASUREMENT RANGE

0 to 100% of localised environmental oxygen concentration

DISPLAY TYPE

Liquid Crystal Display (LCD)

SENSOR TYPE

Class T-7 (Galvanic)

ACCURACY

Approximately 2% linear

SENSOR LIFE

Approximately 12 months in most applications