Teledyne TED 191 User Manual

Page 19

Advertising
background image

19

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

T

ROUBLESHOOTING

S

YMPTOM

C

URE

/ R

EMARK

The displayed readings
are unstable

1) Remove the sensor tip from the T-adaptor, unscrew the “flow

diverter” and gently swab any excess moisture off the sensor cell
membrane, then flow dry gas over the membrane.

2) Check for a source of RFI emission. Relocate the instrument.

3) Check that the calibration control knob is properly locked in

position.

Instrument reads “00”

1) The sensor is not plugged in.

2) There is bad sensor/cable or cable/monitor connection.

Displayed value drifts
by 2 – 3% over 3 – 4
hours

1) Replace sensor cable.

2) Replace cable.

Alarm continuously
activates when
displayed value is less
than 18%

1) The minimum low alarm threshold is 18%. The Teledyne TED 191

will always alarm under these conditions.