3.1

C

LEANING THE

TED 191 P

ORTABLE

O

XYGEN

M

ONITOR

The instrument itself should be cleaned using a damp cloth. In the event of an excessive
build-up of dirt, a mild alcohol solution (isopropyl) may be used to wipe over the instrument,
which should be allowed to air-dry afterwards. In extreme cases, the monitor casing may be
cleaned using a mild detergent.

Note:

The Accurox Oxygen monitor cannot be liquid sterilised or autoclaved –
irreparable damage may result.

3.2

S

TERILISING THE

T-7 O

XYGEN

S

ENSOR AND

C

ONNECTING

C

ABLE

The T-7 oxygen sensors and connecting cables used on the instrument may be gas
sterilised using low temperature ethylene oxide. A vacuum must not be drawn on the
sensor during the sterilisation process.