SWRA004A

TRF1500 Integrated Dual-Band RF Receiver User’s Guide

63

c) After entering the ENR for the desired frequency, press the

Frequency button on the front panel to exit.

5) To calibrate the NF Meter:

a) Connect the Noise Source directly to the NF meter; press

the calibration button twice.

b) Next, press the Noise Figure and Gain Button. The

corrected LED just above the button should be lit.

c) Calibration is complete.

Next the external equipment loss is considered (RF cable,
Transmission line).

6) The Losses are entered in the Noise Figure Meter by using

special function 34.x.

a) Special Function 34.1 turns on the Loss compensation

factor.

b) Special Function 34.2 is used to enter the loss before the

DUT.

c) Special Function 34.3 is used to enter the room

temperature in Kelvin (300

°

K).

d) Finally Special Function 34.4 is used to enter the loss after

the DUT.

e) Special Function 34.0 is used to turn off the loss

compensation factor.

The noise figure is measured as follows:

7) Connecting the noise source directly to the Low Band LNA

input port, J10.

8) Connect the EVM LNA output port, J15, to the noise figure

meter input port.

9) Measure the noise figure for each frequency.

Low-Band LNA: Third Order Input Intercept Point (IIP3)

Control state: 011000

SEE APPENDIX A: TEST BENCH SETUPS

Test setup Figure 26