TELEDYNE ANALYTICAL INSTRUMENTS
13. If the O
2
cell current is too strong or too weak to span,
an error message will appear. This will not occur if the
user-entered span gas concentration is correct. Press
ACK to retry, or ESC to abort the span.
14. If the span is successful, the system will return the main
Analyze mode menu. Check to see that the span value is
correct.