Teledyne 3160 User Manual

Page 52

Advertising
background image

4-24

4 Operations

4 Operations

4 Operations

4 Operations

4 Operations

TELEDYNE ANALYTICAL INSTRUMENTS

7. Status page five shows the time and date of the last time

the alarm was triggered for each of the five program-
mable alarms.

8. Status page six shows the set-up for the serial port, and

allows you to reconfigure the serial port. Press EDIT to
bring up the following screen.
Press NEXT to go to status page 7.

9. Use the SELEC, UP and DOWN keys to select the de-

sired serial port configuration, and press ACCPT.

Baud rate:

Baud rate:

Baud rate:

Baud rate:

Baud rate:

300–2400

Parity:

Parity:

Parity:

Parity:

Parity:Odd, Even, or None
Mode:

Mode:

Mode:

Mode:

Mode: Monitor, Remote, or Off
ASCII Data Bits:

ASCII Data Bits:

ASCII Data Bits:

ASCII Data Bits:

ASCII Data Bits: 7 or 8
Stop Bits:

Stop Bits:

Stop Bits:

Stop Bits:

Stop Bits: 1 or 2
Modem:

Modem:

Modem:

Modem:

Modem: On, Off or Auto
Output Interval for Monitoring:

Output Interval for Monitoring:

Output Interval for Monitoring:

Output Interval for Monitoring:

Output Interval for Monitoring:

1–32000 sec.

Transmit/Receive Line Terminators:

Transmit/Receive Line Terminators:

Transmit/Receive Line Terminators:

Transmit/Receive Line Terminators:

Transmit/Receive Line Terminators: CR or

CRLF

10. Status page seven allows for configuration of the mo-

dem/remote. Press EDIT to change the configuration.

Modem Mode

Modem Mode

Modem Mode

Modem Mode

Modem Mode: AUTO/OFF/ON

Modem Status

Modem Status

Modem Status

Modem Status

Modem Status: NO (set by system)

Login Access

Login Access

Login Access

Login Access

Login Access: LOCAL (set by system)

Phone Jack

Phone Jack

Phone Jack

Phone Jack

Phone Jack: SINGLE/MULTI, RJ12

Comm. Std. for 1200B

Comm. Std. for 1200B

Comm. Std. for 1200B

Comm. Std. for 1200B

Comm. Std. for 1200B: BELL USA/CCITT

11. After pressing EDIT, a cursor will appear and you will

be able to edit the settings. Press ACCPT when done.
Pressing INIT in status page seven (step 10) takes you
to the screen at the top of the next page and gives you
the option of either checking for a modem, or setting up
the modem. Set-up sets up the modem for auto answer
and other communication protocols.