Appendix E: Performance Verification Procedure

A–16

VX4357

Equipment Required

Table 1–1 lists the equipment required for the performance and verification procedure.

Table 1–1: Equipment Required

Required tools and equipment

Part number

VXI Mainframe (such as the Tektronix VX1410)

n/a

VXI Slot 0 with resource Manager (Tektronix VX4521) and appropriate cables and
interface cards.

n/a

Connector, DD-50P

n/a

VX1785S connector

n/a

Talker/Listener (Send/Read)

n/a

4.5 digit Digital Multimeter with a 4-wire Ohms capability and capable of reading
greater then 10 Gigohms.

n/a

Connector Wiring

Using the DD-50P connector wire the following pins together, leaving a pig tale, and label this
N.O.:

1, 3, 5, 7, 9, 11, 13, 15, 18, 20, 22, 24, 26, 28, 30, and 32

Using the VX1785S connector wire the following pins together, leaving a pig tale, and label this
N.C.:

34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, and 49

Default Conditions And LED Test

1. Apply power to the VXI system and wait 10 seconds.

Only the green PWR LED will be lit.

2. Send “S

lf

” (Set all relays) to the VX4357.

3. Verify that all LEDs for CH00 through CH31 are lit.
4. Send “R

lf

” (Reset all relays) to the VX4357.

5. Verify that all LEDs for Ch00 through CH31 are out.

DMM Connection And Set-Up

Connect the DMM signal high and input positive (+) side to the pigtail on the DD-50P connector
that is called out in the testing section.
Connect the DMM signal low and input negative (–) side to the pigtail on the VX1780S connec-
tor that is called out in the testing section. Set the DMM into 4 wire Ohms mode and autorange.