Relay resistance test: channel 00 – Tektronix VX4357 User Manual

Page 62

Advertising
background image

Appendix E: Performance Verification Procedure

VX4357

A–17

Relay Resistance Test: Channel 00

1. Place the DD-50P connector onto P4.
2. Place the DMM + side to the pigtail labeled N.C., and place the DMM – side to DD-50P

pin-2 (wiper Channel 00).

3. The DMM will read less then 400 milliohms.
4. Send “C0

lf

” (Close relay CH00) to the VX4357.

5. The DMM will read greater then 10 Gigohms.
6. Send “Q0

lf

” (Query relay CH00) to the VX4357.

7. Read “1

crlf

” (Relay CH00 is closed).

8. Place the DMM + side to the pigtail labeled N.O .
9. The DMM will read less then 400 milliohms.
10. Send “O0

lf

” (Open relay CH00) to the VX4357.

11. The DMM will read greater then 10 Gigohms.
Repeat steps 1 through 11 for relays CH01 through CH15, substituting Cx

lf

and Ox

lf

for the ap-

preciate relay, where x = relay to close or open. Use Table 1–2 for the channel to wiper pin-out.

Table 1–2: Channel to wiper pin-out

Channel

Wiper pin

01

4

02

6

03

8

04

10

05

12

06

14

07

16

08

19

09

21

10

23

11

25

12

27

13

29

14

31

15

33

12. After repeating the above steps for Ch00 through CH15, remove the DD–50P connector from

P4 and place it on P5.

13. Add 16 to all relay references and repeat the above steps for relays CH16 through CH 31.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com