Troubleshooting, Leak testing, Chapter 4 troubleshooting – Valco electronic Valve Oven User Manual

Page 43: Chapter 4

Advertising
background image

Operating Manual

43

4

Troubleshooting

This chapter has information to help you to find and correct potential problems
when using the TRACE GC Ultra Valve Oven.

Chapter at a Glance…

Leak Testing

.........................................................................................................43

Valve Switching

....................................................................................................44

Valve Plumbing

....................................................................................................45

The most common problem associated with valves and plumbing are leaks, valves
not switching properly, and valves not plumbed correctly.

Leak testing

Leak testing requires studying plumbing diagrams and system configurations.

1. Identify all gas sources leading into the flow path to be leak tested. Determine

if the gas sources are either a common source or if they are at the same bottle
pressure.

2. Be sure all components in the sample path will withstand the pressure being

applied to the system.

3. Be sure all possible gas outlets are blocked. This will require knowing how to

identify the various flow paths throughout the valving system to be sure vents
are blocked or bypassed.

NOTE

Leak should be detected using an electronic leak detector.

Advertising