ABPM 6100 Hardware Manual

DIR 80012720 Ver. B

28 / 30


Electromagnetic Immunity

The ABPM6100 is intended for use in the electromagnetic environment specified below. The customer or the user

of the ABPM6100 should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment

guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output
lines

n.a.

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

n.a.

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

n.a.

Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the ABPM6100 requires
continued operation during
power mains interruptions, it is
recommended that the
ABPM6100 be powered from an
uninterruptible power supply or
a battery.

Power frequency (50/60
Hz) magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.