Abpm 6100 hardware manual – Welch Allyn ABPM 6100 User Manual

Page 30

Advertising
background image

ABPM 6100 Hardware Manual

DIR 80012720 Ver. B

30 / 30


Recommended separation distances between portable and mobile RF

communications equipment and the ABPM6100

The ABPM6100 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the ABPM6100 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the ABPM6100 as recommended below, according to the maximum output power of the communications
equipment.

Separation distance according to frequency of transmitter

m

Rated maximum

output power of

transmitter

W

150 KHz to 80 MHz

P

d

= 2

.

1

80 MHz to 800 MHz

P

d

= 2

.

1

800 MHz to 2,5 GHz

P

d

= 3

.

2

0,01 0.12

0.12

0.23

0,1 0.37

0.37

0.74

1 1.2

1.2

2.3

10 3.7

3.7

7.4

100 12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.


Advertising