ABPM 6100 Hardware Manual

DIR 80012720 Ver. B

23 / 30


7. Specifications

This section provides the unit specifications for the ABPM 6100 ambulatory blood pressure
monitor.
Power Requirements:

• Two “AA” alkaline batteries or high capacity rechargeable batteries (NimH).
Dimensions:

• 124 x 70 x 33 mm (approximate).
Weight:

• 270g (including batteries, approximate).
Operating Ambients:
• Temperature: 10°C (50°F) to 50°C (122°F).

• Humidity: 20% to 95% RH non-condensing.
Storage Ambients:

• Temperature: -20°C (-68°F) to 70°C (158°F).

• Humidity: 15% to 95% RH non-condensing.

• Altitude: -170 to 1,700 meters
Gauging Ranges:

• Systolic Blood Pressure: 60 to 250 mmHg

• Diastolic Blood Pressure: 25 to 200 mmHg

• Maximum Inflation: 270 mmHg

• Heart Rate: 40 to 200 bpm
Memory:

• Up to 250 readings using Alkaline batteries

• Up to 110 reading using rechargeable NimH
Method of Measurement:

• Oscillometric with step deflation.
Accuracy:

• ± 3 mmHg
International Standards:

• EN 60601-1-1:2001 ; Medical electrical equipment – Part 1-1: General requirements for

safety. Collateral standard: Safety requirements for medical electrical systems
Specific requirements for the protection against hazards of ignition of flammable
anesthetic mixtures and to control the risk of fire (Fire prevention)

• EN 60601-2-30:2000 ; Medical electrical equipment – Part 2-30: Particular requirements

for the safety, including essential performance, of automatic cycling non-invasive blood
pressuring monitoring equipment

• EN 60601-1-2:2001 ; Medical Electrical Equipment – Part 1-2: General requirements for

safety. Collateral Standard: Electromagnetic compatibility – requirements and tests

• EN 1060-1:1996 ; Specification for Non-invasive sphygmomanometers – Part 1.

General requirements

• EN 1060-3:1997 ; Non-invasive sphygmomanometers – Part 3: Supplementary

requirements for electro-mechanical blood pressure measuring systems

• AAMI SP 10 ES1 Category C: 1992 (battery powered) Electronic or automated

sphygmomanometers

Calibration:

• Minimally, once per year
Safety System:

• Maximum Inflation pressure limited to 300 mmHg.

• Auto safety release valve for power failure.

• Maximum BP measurement time limited to less than 180 seconds.
Data Connector:

• Stereo mini-plug headphone type connector.