Olympus AU2700 User Manual
Page 39
- 39 -
(3) Kind of T.R.I. transfer/receive.
A. T.R.I. of normal sample
Setting of parameters
Method of T.R.I. inquiry & T.R.I.
Receive T.R.I.. inquiry
Method of
T.R.I. inquiry
Kind of T.R.I.
T.R.I. inquiry
T.R.I.
Real-time
Sequential
Rack no.
Request
of sample no
Sample no
Sample no *1
Sample no
(Calculated rack
No. and cup pos.
No.)
Sample ID
Request
of sample ID
Sample ID
Sample no *1
Sample ID *1
Sample no *1
Batch
Sequential
Rack no.
Request
of sample no
Sample no.
Sample no *1
Sample ID
Sample ID *1
Sample no *1
• Selectable setting of receiving T.R.I. in online parameter menu and requisition format menu.
• Selectable method of T.R.I. in system menu.
Remark *1: Sample no. and Sample ID of T.R.I. must be same to the Sample no. and Sample ID of T.R.I inquiry.
(4) T.R.I. of Repeat sample
Setting of parameters
Method of T.R.I. inquiry & T.R.I.
Repeat sample
Receive of T.R.I.
Repeat rack
Method of
T.R.I. inquiry
Kind of T.R.I.
T.R.I. inquiry
T.R.I.
Real-time
Using
Sequential
rack no.
Request
of Sample no.
Repeat sample no.
Original
sample no. *1
Repeat sample no.*1
Sample ID
Request
of sample ID
Sample ID
Repeat sample no.
*1
Sample ID *1
Repeat sample no.*1
Batch
-
Sequential
rack no.
Request
of sample no.
Repeat sample no.
Original
sample no. *1
Sample ID
Sample ID *1
Remark *1: Repeat sample no. and sample ID of T.R.I. must be same as the Repeat sample no. and sample ID of T.R.I
inquiry.