(1) Flags list

Flag

Contents

(

Failed to evade the contamination since the detergent for evading the contamination is not enough.

%

The sample probe is clogged with sample.

?

Calculation unable due to abnormal photometric data

#

Sample level detection error

R

Reagent level detection error

!

Calculation unable due to abnormal data

U

Reagent absorbance value at P-START of Reagent Blank run, is smaller than the
Lower limit of the Parameter.

u

Reagent absorbance value at P-START is lower than the lower limit specified
In the Parameters in routine run.

Y

Reagent absorbance value at P-END of Reagent Blank run, is greater than the
Upper limit of the Parameter.

y

Reagent absorbance value at P-END is higher than the upper limit specified
In the Parameters in routine run.

@

Abnormally high result; absorbance of every wavelength is more than 2.5.

$

No linearity validation conducted because less than 3 data obtained in the
Kinetics.

D

Too quick reaction slope in increasing kinetics, absorbance at P-START is
Higher than MAX. OD in increasing FIXED assay, or too slow reaction slope
In decreasing kinetics (= no reaction observed)

B

Too quick reaction slope in increasing kinetics, or absorbance at P-END is
Lower than MIN. OD in increasing FIXED assay.
Linearity error in kinetics.

*

Linearity error in kinetics.

&

Prozone data error (prozone check point data error)

Z

Prozone error

F

Result higher than the dynamic range specified in the Parameters

G

Result lower than the dynamic range specified in the Parameters

x

Other QC data error

1

QC result beyond the QC value range specified in the Parameters

2

Multi-rule QC data error

3

Multi-rule QC data error

4

Multi-rule QC data error

5

Multi-rule QC data error

6

Multi-rule QC data error

7

Multi-rule QC data error

p

Result beyond the panic value specified in the Parameters

T

Abnormality found in the Inter-Item Check

P

Result higher than DECIDE RANGE designated in parameters.

N

Result lower than DECIDE RANGE designated in parameters.

H

Higher than the result value range specified in the Parameters

L

Lower than the result value range specified in the Parameters

J

Result higher than the repeat run range specified in the Parameters

K

Result lower than the repeat run range specified in the Parameters

S

Result extracted for repeat run