Timing/time-out – Olympus AU2700 User Manual

Page 33

Advertising
background image

- 33 -

(4.2) Timing/time-out

(1) Time-out/timing list .

Kind Description

Default

Remarks

T1

Limit time from end of transfer/receive to start
of text receive.

2 s

0.1 x n sec

T2

Limit time from start of text receive to end of
text receive.

(Note 1)

(Note 2)

T3

Limit time from start of text transfer to end of
text transfer.

n=1-99

T4

Limit time from end of text transfer to response of
receive.

2 s

T5

Interval time of transmission

2 s

T6

Min. time from receive

1 s

T7

Limit time from NAK response to start receive
of retry text.

2 s

t1

Min. time from end of transfer text to receivable

0.5 s

Not Changeable

text.

t2

Min. time from end of transfer text to next receivable
text.

0.5 s

t3

Min. time from end of transfer text to receivable
Response.

0.5 s

t4

Min. time from end of receiving text to transfer
Response.

0.5 s

note 1) ((Max text length x Character length)/Bit/sec)+ 0.5 s
note 2) These time-outs can be changed in the online parameters menu.

(2) Rule of class A

A. Case 1

AU2700/5400

A

C

Host

B

<T3

<T2

<T3

t1<<T1 T5<

B. Case 2

AU2700/5400

A

B

C

Host

<T3

<T3

<T3

T5< T5<

Advertising
This manual is related to the following products: