Sending test result protocol – Olympus AU2700 User Manual

Page 40

Advertising
background image

- 40 -

(5.2) Sending test result protocol

(1) Transfer sequence for one session.

General sequence :

A. Example 1

Shift next Dx-transfer when Online Error occurred.

AU2700/5400

DB

Dx

0

Dx

E

Dx

0

Dx

E

Dx

E

DE

2 blocks
Host

B. Example 2

Stop session when Online Error occurred.

AU2700/5400

DB

Dx

0

Dx

E

Dx

0

Dx

E

Dx

E

DE

Host

(2) Detail

Data
Trans. Kind of text

Timing/condition

Normal end

Online error

Real

DB (START)

This text is transmitted when

It is shifting the

It is output the

-time

it is shifted from stand-by

transfer of the

alarm and it is

mode to measure mode and just

test results.

executed the below

behind the first Dx is transmitted.

• Error protocol is

Dx (result data)

It is transmitted one by one

It is shifting next

stop.

the samples that are finished

The transfer of

It is canceled the

measuring.

Dx/DE.

transfer of Dx/DE

• Error protocol is

continue
It is shifting next
the transfer of
Dx/DE.

DE (end)

This text is transmitted after

It is ended the

It is output the

the last transfer of Dx text

session of transfer of

alarm and ended

and the below shifting mode.

The text of test

the session of the

results.

transfer of the

1) measure mode to standby

text of test results.

2) measure mode to stop
This text is transmitted when
the communication is canceled
by the online alarm.

Advertising
This manual is related to the following products: