Specification of connection, In/output signals and terminals, Signal level – Olympus AU2700 User Manual

Page 43: Applicable connector, Others

Advertising
background image

- 43 -

6. Specification of Connection

(6.1) In/Output signals and Terminals

PC Terminal

Terminal

at AU2700/5400

at host

Signal Name Abbrv. Direction

computer

(25 Pin Configuration)

-------------------------------------------------------------------------------------------------------------------
Ground

FG

-

1

Data sent

TxDATA

3

3 (Rx DATA)

Data receive RxDATA

2

2 (Tx DATA)

Signal Ground SG

5

7 (SG)

Request to send RTS 7
Clear to send CTS 8

Caution: RTS and CTS must be connected directly.

(6.2) Signal Level

Signal/Signal Format Signal Level

------------------------------------------------------------------------------------

SPACE (ON)

+3V and higher ("H")

MARK (OFF)

-3V and lower ("L")

Note: "H" stands for "HIGH LEVEL" and "L" for "LOW LEVEL".

(6.3) Applicable Connector

Style of connector : Confirm the standard of IBM PC (RS232 DB9 Male)
Provide a RS232 DB9 Female connector.
Style of connector Cover : ditto

(6.4) Others

A. When powering ON or OFF the host computer system, ensure that the AU2700/5400 is in
Standy mode. Powering ON or OFF the host computer system while the AU2700/5400 is being operated may
result in abnormality function of the AU2700/5400.

B. The cable for host communication is not included in the AU2700/5400 standard accessories.
A shielded cable must be used for host connection.

C. Do not use the terminals on the AU2700/5400 other than No. 2,3,5,7, or 8 for any purpose.

Advertising