Gas sampling and conditioning technology

5-5.3 ME

9.3

CONNECTION OF THE SENSOR TYPE KS2.EX TO THE ELECTRONICS WE77/EX1

N O T E !

The bridge between the connecting points 3 and 4 of the electronics
WE77/Ex1 guarantees a „safty-first“ alarm release (alarm in case of voltage
loss and parting of the cable).


The function is as follows:

dry sensor : LED

„ON“ contact MC-NO terminal 7 and 9 closed Sensor current > 1,7 mA

wet sensor :

LED „OFF“ contact MC-NC terminal 7 and 8 closed Sensor current < 1,4 mA

Brigde between connecting points 4 and 5 or missing bridge

N O T E !

If the bridge is missing or switched between the connecting points 4 and 5 of
the electronics WE77/Ex1 the alarm is not working in a „safty-first“ mode (no
alarm in case of voltage loss).


The sensor current is on a very low level. Therefore we recommend to switch a resistance of 10k

parallel to the sensor to increase the level.

The function than is as follows:

dry sensor :

LED „OFF“ contact MC-NC terminal 7 and 8 closed Sensor current > 1,7 mA


wet sensor :

LED „ON“ contact MC-NO terminal 7 and 9 closed Sensor current < 1,4 mA



9.4

CONNECTION OF THE SENSOR TYPE KS2.EX TO THE ELECTRONICS
KFA6-SR2-EX1.W

Move all switches (S1-S3) on the front into position I (left).

The function is as follows:

dry sensor : left

LED „OFF“ contact MC-NC terminal 7 and 8 closed Sensor current > 1,7 mA


wet sensor : left

LED „ON“ contact MC-NO terminal 7 and 9 closed Sensor current < 1,4 mA