Adjustment of the sensor type ks2.ex, Figure 4, Figure 5 – M&C TechGroup KS 2.Ex Operator's manual User Manual

Page 12: 5 adjustment of the sensor type ks2.ex, Kfa6, Sensor ks2.ex

Advertising
background image

12

Gas sampling and conditioning technology

5-5.3 ME

9.5

ADJUSTMENT OF THE SENSOR TYPE KS2.EX

Figure 4

Electronic KFA6-SR2-Ex1.W (left) and terminal box KS2.Ex respectively WE77/Ex1

Turn the potentiometer (figure 4) to the left until the LED is OFF.

Turn the potentiometer slowly to the right. After the LED is ON, turn the potentiometer for another

1,5 rotations to the right.


1

-

2

+

3

KFA6

1

2

3

Sensor KS2.Ex

-

+

Switching capacity AC 253V/2A

S1
S2
S3

14

N

15

L

Power supply

230V 50Hz

1W

13

7

8

9

Output

MC NO NC

Figure 5

Wiring diagram KS2.Ex and KFA6-SR2-Ex1.W respectively WE77/Ex1

Bridge

Left

Right

Electronics

WE77/Ex1

Poti

Sensor

1

-

2

+

3 4 5 6

7

MC

8

NC

9

NO

10

PE

11

L

12

N

WE77/Ex1

Power supply

230V 50Hz

3,5VA

Bridge

1 2 3

Sensor KS2/Ex

-

+

AC 250V/4A/500VA
DC 229V/0,1A; 60V/ 0,6A; 24V/4A