Application, Figure 1, Ks2 / ks2.ex liquid-sensor – M&C TechGroup KS 2.Ex Operator's manual User Manual
Page 7
7
Gas sampling and conditioning technology
5-5.3 ME
6
APPLICATION
Liquid alarm sensors are used in gas sample conditioning systems for monitoring gas cooling and
condensate removal devices in order to provide protection for downstream analysis instruments. This
monitoring device KS2 / KS2.Ex reliably signals the penetration of non conductive (e.g. alcohol)
and conductive (e.g. water) liquid in the event of cooling or condensate removal equipment being
defective, thus avoiding expensive down time as well as high repair costs for analysis instruments.
In the event of an alarm, we recommend to switch off the power supply for the sample gas or a
solenoid valve in the sample conditioning system.
PG 7
ø16
135
65 x 50
Sleeve nut GL25/18
Chemraz* O-ring 15
KS 2
* Chemraz is a Greene Tweed Trademark
Figure 1
KS2 / KS2.Ex Liquid-Sensor