2 record setup – Soft dB Opus Suite Data Logger Module User Manual

Page 14

Advertising
background image

13

Step 2

Install the calibrator device on the microphone

Step 3

Click START

After the average time is elapsed, the sensitivity value will update.

Step 4

Click OK to accept the sensitivity value

6.2 Record Setup

Record Setup

Record Path

Click on the control to adjust the destination path and name of the .wav to

be recorded. This uses the Explorer Window (see section 11, p. 49).

Auto-Increment
Set the switch to

or

to enable or disable the auto-increment of

the record file.

When the auto-increment is on, the name must be defined only for the first

measurement. All others measurements will be logged in a wave file with a

four-digit extension number.

For instance:

First measurement file name: test.wav

Second measurement file name:

test_0001.wav

Third measurement file name: test_0002.wav

Wave File Length in the format hh:mm:ss (hour:min:sec).

In order to avoid large files that would be heavy to post-process, the data

logging will switch wave file as it reaches this length. The length is limited to

a minimum five minutes and is further limited to a Wave File Size of 1 GB.

The partition index will be had to the specified measurement name as new

partitions are created.

For instance:

Measurement name (length = 5 min): test_0002.wav

1st partition (from 0 to 5 min) : test_0002.wav

2nd partition (from 5 to 10 min):

test_0002_2.wav

Nth partition (from 5∙(N-1) to 5∙N min): test_0002_N.wav