V

essel pressure

-

S

crewed part PVDF

-100 … 300

kPa/-1 … 3 bar (-14.5 … 43.5 psig)

-

S

crewed part and flange 316L

-100 … 1600

kPa/-1 … 16 bar (-14.5 … 232 psig)

P

rocess temperature (measured on the process fitting)

-

P

rocess fitting, thread PVDF

-40 … +130 °C (-40 … +266 °F)

-

O

ther process fittings

-40 … +150 °C (-40 … +302 °F)

V

ibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

8)

Electromechanical data - version IP 66/IP 67

C

able entry

-

D

ouble chamber housing

l

1

x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),

1

x blind stopper M20 x 1.5; plug M12 x 1 for

VEGADIS

61 (

optional)

or:

l

1

x closing cap ½ NPT, 1 x blind stopper

½

NPT,

plug M12 x 1 for VEGADIS 61

(

optional)

or:

l

1

x plug (depending on the version), 1 x blind

stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (

optional)

S

pring-loaded terminals for wire cross-sec-

tion

2.5

mm² (AWG 14)

Indicating and adjustment module

V

oltage supply and data transmission

through the sensor

I

ndication

LC

display in dot matrix

A

djustment elements

4

keys

P

rotection

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

M

aterials

-

H

ousing

ABS

-

I

nspection window

P

olyester foil

Power supply

O

perating voltage

-

N

on-Ex and Ex-d instrument

20 … 72 V DC, 20 … 253 V AC, 50/60 H

z

M

ax. power consumption

4 VA; 2.1 W

8)

T

ested according to the regulations of German Lloyd, GL directive 2.

VEGAPULS

65 • 4 … 20

mA/HART four-wire