2 technical data, 1 data – VEGA D77 Hydrostatic pressure transmitters User Manual

Page 7

Advertising
background image

Pressure transmitters D76 and D77

7

12

300

600

900

18

24

30

36

Product description

2 Technical data

2.1 Data

Supply and signal current circuit (analogue transmission, 4 … 20 mA)

Power supply

12 … 36 V DC

Residual ripple of the power supply

no influence at U

SS

< 100 mV

Output signal
- terminal insert

analogue transmission (not standardised)
4 … 20 mA

- adjustment insert

4 … 20 mA (adjustable)

Current limitation

approx. 22 mA

Fault signal

22 mA

Integration time

0 … 10 s (adjustment time from 10 % … 90 %

of measuring range final value)

Average delay time

85 ms

Connection cable

2-wire

Max. permissible load

depending on power supply
see following load diagram

Load in

W

Supply voltage
in V

Integrated overvoltage protection (option)

Nominal d.c. spark-over voltage
- protective diode

40 V

- gas separator

650 V

Nominal discharge current
- gas separator

20 kA