VEGABAR 41 (Profibus PA)

9

Product description

Operating conditions

Ambient conditions

Ambient temperature

-40°C … +85°C

- with display module

-20°C … +70°C

Storage and transport temperature

-40°C … +85°C

Product temperature, dependent on
isolating liquid

Temperature
p

abs

>

1 bar

/

p

abs

< 1 bar

- silicone oil

-40°C … +150°C

/

-40°C … +130°C

- silicone oil and cooling element

-40°C … +200°C

/

-40°C … +130°C

- silicone oil and capillaries 1 m

-40°C … +200°C

/

-40°C … +130°C

- vegetable oil

-10°C … +130°C

/

-10°C … +130°C

- vegetable oil and cooling element

-10°C … +200°C

/

-10°C … +130°C

- high temperature oil a. cooling element -10°C … +400°C

/

-10°C … +200°C

- high temperature oil and capillaries 1 m -10°C … +400°C

/

-10°C … +200°C

Additional temperature influence of isolation diaphragm

Process connection

Temperature coefficient

[mbar/10 K]

FA

Flange DN 25 PN 40 acc. to DIN 2501,
seal surface acc. to DIN 2526 form D

1,8

FB Flange DN 40 PN 40 acc. to DIN 2501,

seal surface acc. to DIN 2526 form D

1,0

FC Flange DN 50 PN 40 acc. to DIN 2501,

seal surface acc. to DIN 2526 form D

2,0

FE Flange DIN 50 PN 40 with ext. 100 mm

1,9

F G Flange DN 50 PN 40 with ext. 200 mm

2,1

FH Flange DN 80 PN 40 acc. to DIN 2501,

seal surface acc. to DIN 2526 form D

0,4

FK Flange DN 80 PN 40 with ext. 100 mm

0,5

FN Flange DN 1" 300 lbs acc. to ANSI B 16.5,

seal surface RF

1,8

F O Flange DN 2" 600 lbs acc. to ANSI B 16.5,

seal surface RF

2,0

- by cooling elements,

depending on diaphragm ø

0.1 … 1.5 mbar/10 K

- by 1 m capillary line,

depending on diaphragm ø

0.1 … 15 mbar/10 K