VEGA VEGAFLEX 52K User Manual

Page 10

Advertising
background image

10

VEGAFLEX 52K

Product description

Connection cables

Two-wire sensor
- Supply and signal via one two-wire cable.

The cable resistance depends on the supply voltage (see diagram).

Four-wire sensor
- Supply and signal separately,

cable resistance of the 4 … 20 mA signal cable max. 500

W

Profibus PA sensor
- recommended cable types

- SINEC 6XV1 830-5AH10 (Siemens AG)
- SINEC L2 6XV1 830-35H10 (Siemens AG)
- 3079 a (Belden)

- max. cable length

100 m field distributor to sensor

Conductor cross-section

generally 2.5 mm

2

Ground connection

max. 4 mm

2

Cable entry

2 x M20 x 1.5 (cable diameter 5 … 9 mm)

CE conformity

VEGAFLEX sensors meet the protective regulations of EMC (89/336/EWG) and NSR
(73/23/EWG). Conformity has been judged acc. to the following standards:
EMC

Emission

EN 50 081 - 1: 1992

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

Display

Display

scalable, analogue and digital display of
measured values (option).

An external display of measured value, powered by the sensor, can be mounted at a dis-
tance of up to 25 m from the sensor.

Signal output

Signal output
- two-wire technology

4 … 20 mA (see diagram)

- four-wire technology

4 … 20 mA

Resolution of the 20 mA signal

0.025 % of range

Load

0 … 500

W

Two-wire technology:
The analogue 4 … 20 mA output signal (measuring signal) is transmitted together with the
power supply via one two-wire cable.

Four-wire technology:
Separate power supply.
The analogue 4 … 20 mA output signal (measuring signal) is transmitted in a cable sepa-
rated from the power supply.

Advertising