Evaluation, Outputs – VEGA VEGABAR 40 4 … 20 mA User Manual

Page 21

Advertising
background image

VEGABAR 40 (4 … 20 mA)

21

Setup

3 Adjustment of zero

Adjust zero with the "+“ or "–“ key
and save with the "OK“ key.

4 Adjustment of span

Adjust span with the "+“ or "–“ key
and save with the "OK“ key.

Note:

- A modification of zero does not influence

the adjusted span, i.e. the measuring
range final value is shiftted.

- When pushing the "+“ or "–“ keys individu-

ally, the current output remains at the last
value, it takes on the adjusted value only
after storing with the "OK“ key.

Evaluation

Adjustment of the integration time

An integration time t

i

of 0 s … 10 s

can be adjusted with the "+“ or "–“
key. The adjusted value can be

saved with the "OK“ key.

Selection of the indicated unit

The actual measured pressure is
indicated in the DOT-matrix in the
measured value indication. The
respective unit can be chosen with
the "+“ or "–“ key and the "OK“ key
from a list. When choosing the unit
"Scal.“, the following menu points
are available.

Scaling indication

In operating condition, the actual measured
pressure is displayed on the indicating mod-
ule:
- as bar graph with 20 segments
- as 4-digit digital value.

ti

1

s

Zero

0,0

mbar

Span

100,0

mbar

Bar graph and digital value relate to the ad-
justed measuring range and change propor-
tionally to the current output. Digital values
< –10 % or > 110 % are shown as flashing
figures.

As a result of scaling, the 4 mA or 20 mA
user-specific current output values are as-
signed to the indicating module as 4-digit
digital value.

Outputs

Current output/Selection of the character-
istics

The current output provides the actual meas-
ured pressure as an analogue current signal
4 … 20 mA relating to the adjustment.

The characteristics can also be
inverted, e.g. switched to
20 … 4 mA with the "+“ or "–“ key
and the "OK“ key.

Fault signal

If, during the continuous self-monitoring,
errors, damage or malfunctions in the meas-
uring cell or the electronics are detected, a
fault signal is triggered via the current output.

With "+“ or "–“ key and the "OK“
key you can choose if the failure
current should be 22 mA or
3.6 mA.

Unit

0.0

mbar

psi
kPa
%
Scal.
mA

Scaling

indica-

tion

100%=

1000

Decimal

point

888.8

0%=

0

Escape

Current

output

4-20mA

20-4mA

Fault

signal

22mA

3.6mA

Advertising