6 instrument modification, 1 exchange of adjustment modules – VEGA VEGABAR 40 4 … 20 mA User Manual

Page 25

Advertising
background image

VEGABAR 40 (4 … 20 mA)

25

Instrument modification

6 Instrument modification

6.1 Exchange of adjustment

modules

The modular construction of VEGABAR allows
retrofitting, interchange or removal of the
adjustment and display modules. Data previ-
ously saved (e.g. adjustment values) are not
stored in the adjustment module but in an
EEPROM of the sensor electronics and there-
fore readjustment is not necessary. The con-
nection of the modules is made via a 4-pole
plug connection.

Exchange of the adjustment module

Removal of the adjustment module

Separate VEGABAR from the power
suppy.

Loosen the screws on the upper side of
the housing and remove the cover or the
display module.

Loosen connection cables from the termi-
nals, if necessary loosen the plug con-
nection of the display module.

Loosen the two screws of the adjustment
module.

Remove adjustment module and pull out
plug connection.

Insertion of an adjustment module

Plug the plug connection of the new ad-
justment module into the plug-in socket of
the sensor electronics.

Fasten the new adjustment module.

Reconnect the connection cables, if nec-
essary cable from the display module.

Close cover or the display module of
VEGABAR.

Reconnect VEGABAR to power supply.

VEGABAR

Module
without
adjustment

Module for
menu-driven
adjustment

Module for
basic
functions

Display
module

Advertising