Setup

2 Adjustment of span

(e.g. process pressure or vessel level at
maximum)
• Set rotary switch to span
• By pushing the "+“ and "–“ keys (simulta-

neously) the current jumps directly to
20 mA or set it to 20 mA by pushing the "+“
and "–“ keys.

Adjustment range of span:
3.3 % … 120 % of the nominal range (corre-
sponds to turn down 1 : 30)

Note:

- A modification of zero does not influence

the span.

- It is also possible to adjust currents for

partial fillings or partial pressures, e.g.
8 mA for 25 % and 16 mA for 75 %. VEGA-
BAR automatically calculates the current
values for 0 % or 100 % (only possible at

3.3 %).

Integration time

An integration time t

i

of 0 … 10 s can be ad-

justed for damping of pressure shocks.

Procedure
• Set rotary switch to t

i

• By pushing the "-“ key 10 times make sure

that the integration time is set to 0 s

• For every 1 s requested integration time,

push the "+“ key once

The integration time is the time required by
the current output signal to reach 90 % of the
actual level after a sudden change of the
process pressure.