VEGABAR 41 (4 … 20 mA)

6 Instrument modification

6.1 Exchange of adjustment

modules

The modular construction of VEGABAR allows
retrofitting, interchange or removal of the
display module. Data previously saved (e.g.
adjustment values) are not stored in the ad-
justment module but in an EEPROM of the
sensor electronics and therefore readjust-
ment is not necessary. The connection of the
modules is made via a 4-pole plug connec-
tion.

VEGABAR

Module
without
adjustment

Module for
menu-driven
adjustment

Module for
basic
functions

Display
module

Diagnostics, Instrument modification

External
energy source

Checking pressure compensation

Open the housing of VEGABAR. The meas-
ured value must not change. However, if the
displayed value changes, compensation of
the atmospheric pressure is not ensured,
which will lead to distortion of the measured
value. Check the breather on the VEGABAR-
housing.

Checking the electrical components

Voltage

- the terminal voltage on VEGABAR must be

at least 12 V DC

Current

Current value

Condition

3.8 … 20.5 mA

standard range for the output
current

0 mA

signal line interrupted

< 3.6 mA

sensor electronics or pressure
sensor element defective

22 mA

shortcircuit in the signal line,
sensor electronics or pressure
sensor element defective

~

1 2 3

+

+

mA

Voltage

Current