VEGAMET 407 AF/AFT

Technical data

Power supply

Value

Tolerance

Frequency

standard

AC

220 V

+15 % -10 %

50 / 60 Hz

AC

230 V

+10 % -15 %

options

AC

110 V

50 / 60 Hz

130 V

+15 %

240 V

-10 %

260 V

AC

24 V

+15 %

50 / 60 Hz

42 V

-10 %

48 V

DC 9 … 16 V

16 … 60 V

Power consumption at U

N

:

approx. 7 VA

Voltage for the transducer on
terminals 1 and 2 (terminal 3 and 4 bridged)

approx. 22 V

Resistance per conductor to the transducer

max. 50 Ohm

Voltage for the transducer on
terminals 1 and 2 (terminal 3 and 4 open)
(only for electrodes with oscillator type 117)

approx. 15 V

Resistance per conductor to the transducer

max. 50 Ohm

Indication

analog indication 0 … 100 %

Output 1

type

current output

range

0 … 20 mA or 4 … 20 mA

load

max. 750 Ohm

floating

VEGAMET 407 AFT

Output 2

type

current output

range

0 … 5 V, max. 2 mA
for connection of
max. 10 auxiliary limit switches VEGASEL

Output 3

type

voltage output

range

0 … 10 V, max. 2 mA

Output 4

type

relay output

function

fail safe

contact

1 spdt

relay data:
contact material

AgCdO and Au plated

min.

turn-on voltage

10 mV

switching current

10

µ

A

max. turn-on voltage

AC 250 V, DC 60 V

switching current

AC 2 A, DC 1 A

breaking capacity

125 VA, 60 W

Integration time

0 … 20 sec.

Permissible ambient temperature on the housing

-20

°

C … +50

°

C / -4 … 122

°

F

Storage and transport temperature

-20

°

C … +70

°

C / -4 … 158

°

F

Housing

type

plastic ABS light grey

protection

IP 40

dimensions

B = 53 mm, T = 143 mm, H = 127 mm

Terminal board: for cross section area of conductors

max. 1 x 1,5 mm

2

Weight: with plug-in socket

approx. 710 g