7 Maintenance and fault rectification

7.1 Maintenance

When used as directed in normal operation, VEGATOR 256C

is completely maintenance free.

7.2 Remove interferences

VEGATOR 256C offers maximum reliability. Nevertheless,

faults can occur during operation. These may be caused by the

following, e.g.:

l

Measured value from sensor not correct

l

Voltage supply

l

Interference on the cables

The first measures to be taken are to check the input/output

signal. In many cases, the causes can be determined this way

and the faults rectified.

However, should this measures not be successful, call the

VEGA service hotline in urgent cases under the phone no. +49

1805 858550.
The hotline is available to you 7 days a week round-the-clock.

Since we offer this service world-wide, the support is only

available in the English language. The service is free of

charge, only the standard telephone costs will be charged.

?

The signal conditioning instrument does not switch when

the respective probe is covered or uncovered

l

Conductivity of the product too low

à

Check if the conductivity value of your product is at

least 7.5 µS/cm

l

Line break to the sensor

à

Check the connection cables from the probe to the

signal conditioning instrument

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) in the Internet from

our homepage www.vega.com under: "Downloads - Forms

and Certificates - Repair form".

Causes of malfunction

Remove interferences

24 hour service hotline