5

Unscrew the pressure transmitter

6

Bring the ball valve in position "Cleaning" by using the
handle. The measurement aperture on the process side is
now visible.

7

Clean the opening with a suitable water jet.

8

Bring ball valve into position "Service" by using the handle

9

Screw the pressure transmitter in until it stops, max. torque
50 Nm

10 Hold the ring and, using the handle, bring the ball valve

again into position "Operation"

11 Turn handle 3½ turns to the right so that the catch is locked

again

Caution:
It is recommended to remove the handle after locking the
catch. This protects the ball valve fitting PASVE against
unauthorized access.

14

Ball valve fitting PASVE • for VEGABAR 54

Mounting

30177

-EN

-080307