VEGA VEGASON 65 Foundation Fieldbus User Manual

Page 56

Advertising
background image

Secondary_value_1

Calibration

Lowest Point

Calibration

Highest Point

Sensor_value

cal_level_lo

cal_level_hi

Calibration
Lowest Point

Calibration
Highest Point

Sensor reference plane

cal_level_lo [%]

cal_level_hi [%]

Fig. 32: Adjustment VEGASON 65

Parameter list for Device revision 3.0
T

he following list contains the most important parameters and their meaning:

l

primary_value

-

T

his is the process value after adjustment and Linearization with the status of the transducer

block

l

primary_value_unit

-

U

nit code of 'Primary_value'

l

secondary_value_1

-

V

alue after min./max.-adjustment (level + level offset). Selected as input to AIFB by setting

'C

hannel' = 2. Unit derives from 'Secondary_value_1_unit'

l

secondary_value_1_unit

-

U

nit code of 'Secondary_value_1'

l

secondary_value_2

-

S

ensor value + sensor offset. Selected as input to AIFB by setting 'Channel' = 3. Unit derives

from 'Secondary_value_2_unit'

l

secondary_value_2_unit

-

U

nit code of 'Secondary_value_2'

56

VEGASON

65 • F

oundation Fieldbus

10 S

upplement

28794

-EN

-081127

Advertising