25

7 Maintenance and fault rectification

VEGACAP 62 • - two-wire

30007-EN-130916

2

1

Fig. 17: Compensation key
1 Measuring range selection switch (compensation key)

2 Control lamp

16. Keep the measuring range selection switch (1) pushed until the

control lamp (2) flashes green.

17. Carry out the adjustment again. See chapter "Set-up, adjustment

elements".

18. Screw the housing cover back on
The electronics exchange is now finished.

7.4 Shortening of the probe

The rod of the probe can be shortened to any length.
1. Shorten the rod probe to the requested length with a metal saw.
2. Mount the probe into the vessel. Make sure that the probe is

uncovered.

2

1

Fig. 18: Compensation key
1 Measuring range selection switch (compensation key)

2 Control lamp

3. Keep the measuring range selection switch (1) pushed until the

control lamp (2) flashes green.

4. Hence the probe is compensated to the modified length.
5. Carry out the adjustment again. See chapter "Set-up, adjustment

elements".

7.5 How to proceed in case of repair

You can find a repair form as well as detailed information on how to

proceed under www.vega.com/downloads and "Forms and certifi-

cates".
By doing this you help us carry out the repair quickly and without hav-

ing to call back for needed information.
If a repair is necessary, please proceed as follows:

Print and fill out one form per instrument