VEGA VEGASON 62 Profibus PA User Manual

Page 38

Advertising
background image

38

8 Maintenance and fault rectification

VEGASON 62 • Profibus PA

28785-EN-130417

If there is no electronics module available on site, one can be ordered

from the VEGA agency serving you.

The new electronics module must be loaded with the settings of the

sensor. These are the options:

At the factory by VEGA

Or on site by the user

In both cases, the sensor serial number is necessary. The serial num-

bers are stated on the type label of the instrument, inside the housing

or on the delivery note.
Information:

When loading on site, first of all the order data must be downloaded

from the Internet (see operating instructions manual "Electronics

module").

The electronics modules are adapted to the respective sensor and

distinguish also in the signal output or power supply.

8.4 Software update

The software version of VEGASON 62 can be determined as follows:

on the type label of the electronics

Via the display and adjustment module

via PACTware

You can view all software histories on our website www.vega.com.

Make use of this advantage and get registered for update information

via e-mail.
The following components are required to update the sensor soft-

ware:

Sensor

Voltage supply

VEGACONNECT

PC with PACTware

Current sensor software as file

At "www.vega.com/downloads" go to "Software". Select under "plics

sensors and instruments", "Firmware updates" the respective instru-

ment series and software version. Load the zip file via the right mouse

key with "Save target as" e.g. on the desktop of your PC. Move with

the right mouse key to the folder and select "Extract all". Save the

extracted files, for example on the desktop.

Connect the signal conditioning instrument to power supply and

provide the connection from the PC to the instrument via the interface

adapter. Start PACTware and go via the menu "Project" to the VEGA

project assistant. Select "USB" and "Set instruments online". Activate

the project assistant with "Start". The assistant establishes the con-

nection automatically and opens the parameter adjustment window

"Sensor # online parameter adjustment". Connect this parameter

adjustment window before you carry out further steps.

Sensor serial number

Assignment

Load sensor software

to PC

Prepare update

Advertising