VEGA VEGAVIB 65 - two-wire User Manual
Page 22
Mode on the
sensor
Level
Signal current -
Sensor
Signal lamp -
sensor
Fault signal
(min./max. mode)
any
<2.3 mA
flashes red
VEGAVIB 65 level switch with signal conditioning instru-
ment
1)
The following chart provides an overview of the switching
conditions depending on the adjusted mode of the signal
conditioning instrument and the level.
Note:
Keep in mind that the mode switch of VEGAVIB 65 must be
always set to "max.".
Mode on the
signal condi-
tioning in-
strument
Level
Signal cur-
rent - Sen-
sor
Signal lamp
- sensor
Signal
lamp - sig-
nal condi-
tioning in-
strument
Mode A
Overfill pro-
tection
approx. 8 mA
Green
Mode A
Overfill pro-
tection
approx.
16 mA
red
Mode B
Dry run pro-
tection
approx.
16 mA
red
Mode B
Dry run pro-
tection
approx. 8 mA
Green
1)
You will find suitable signal conditioning instruments under "Technical data"
22
VEGAVIB 65 - with two-wire output
Set up
29286
-EN