VEGA VEGAVIB 65 - two-wire User Manual
Page 25
Advertising
Sensor
current
VEGAVIB
Level re-
lay VEGA-
TOR A -
Overfill
protection
Level re-
lay VEGA-
TOR B -
Dry run
protection
LED
VEGATOR
A - Overfill
protection
LED
VEGATOR
B - Dry
run pro-
tection
Fail safe
relay
VEGATOR
LED
1. Fault signal
(approx. 3 s)
<2 mA
currentless currentless
currentless
2. Empty signal
(approx. 1.5 s)
approx.
8 mA
energized
currentless
energized
3. Full signal
(approx. 1.5 s)
approx.
16 mA
currentless energized
energized
4. Return to the actual
operating condition
Note:
When used in measuring chains according to IEC 61508,
mode B is not permitted.
With the stated current values you can carry out the function
test directly via a safety SPLC or a process control system.
VEGAVIB 65 - with two-wire output
25
Set up
29286
-EN